Accurate measurement of effective ion concentration in the lab
Measuring effective ion concentration with ion selective electrodes (ISEs) is important in applications such as wastewater, drinking water, environmental testing, and the food/beverage industry. The TruLine series of combination ISEs is designed to meet the most demanding challenges of determining effective ion concentration in the laboratory setting. The TruLine ISEs can be connected to the YSI TruLab 1320(P), the YSI MultiLab 4010-2 (with BNC adapter), the YSI MultiLab 4010-3 (with BNC adapter), and other ISE instruments that feature BNC connection. The TruLine ISEs are refillable and feature a built-in double-junction reference system, allowing for a long electrode life and optimum performance in the lab.
YSI TruLine ISE Specifications |
||||||||
Parameter |
Item # |
Technology |
Features |
Range |
Slope |
pH Range |
Interferences |
Applications/ Notes |
Ammonia (NH3)* |
400365 |
Gassensing |
Replaceable/refillable membrane modules (1 replacement included), pH sensor inside |
0.02 to 17,000 mg/L |
54 to 59 mV/ decade |
ISA buffers sample to >pH 11 to make NH3 gaseous |
Few, volatile amines |
Wastewater, aquaculture, agriculture, industrial. pH shift caused by ammonia gas is correlated to conc. |
Ammonium (NH4+) |
400371 |
Polymer/ PVC membrane |
Refillable reference, replaceable sensor modules (1 replacement included) |
0.02 to 1,800 mg/L |
54 to 59 mV/ decade |
4 to 10 |
Few, Na+, K+ |
Wastewater, environmental, industrial, food/ beverage |
Bromide (Br-) |
400377 |
Solid state |
Refillable reference, nonreplaceable sensor |
0.4 to 79,900 mg/L |
54 to 59 mV/ decade |
<12 |
I-, Cl-, CN-, NH3- |
Pool, spa, industrial, wastewater |
Cadmium (Cd2+) |
400382 |
Solid state |
Refillable reference, nonreplaceable sensor |
0.1 to 11,200 mg/L |
26 to 29 mV/ decade |
2 to 8 |
Hg2+, Ag+, Cu2+, Pb2+ |
Industrial, environmental |
Calcium (Ca2+) |
400387 |
Polymer/ PVC membrane |
Refillable reference, replaceable sensor modules (1 replacement included) |
0.02 to 40,000 mg/L |
25 to 29 mV/ decade |
2.5 to 11 |
Pb2+, Hg2+, Si2+, Mg2+, Cu2+, K+, Tris+, Na+, others |
Aquariums, industrial, food/beverage |
Chloride (Cl-) |
400393 |
Solid state |
Refillable, reference nonreplaceable sensor |
1.8 to 35,500 mg/L |
54 to 59 mV/ decade |
2 to 12 |
CN-, Br-, I-, OH-, S2-, NH3 |
Industrial, wastewater, drinking water, food/ beverage, environmental. Store in dark when not in use. |
Copper (Cu2+) |
400445 |
Solid state |
Refillable reference, nonreplaceable sensor |
0.06 to 6,400 mg/L |
26 to 20 mV/ decade |
2 to 6 |
Hg2+, Ag+, Fe2+, Br-, Cl- |
Industrial, environmental. Also known as Cupric ISE. |
Cyanide (CN-)** |
400403 |
Solid state |
Refillable reference, nonreplaceable sensor |
0.2 to 260 mg/L |
54 to 59 mV/ decade |
10 to 14 |
I-, Br-, Cl-, S2- |
Industrial, environmental |
Fluoride (F-) |
400406 |
Solid state crystal |
Refillable reference, nonreplaceable sensor |
0.02 mg/L - saturation |
54 to 59 mV/ decade |
5 to 7 at low range, 11 at higher range |
OH- |
Drinking water, industrial. Lanthanum fluoride crystal is the sensor. |
Iodide (I-) |
400411 |
Solid state |
Refillable reference, nonreplaceable sensor |
0.006 to 127,000 mg/L |
54 to 59 mV/ decade |
<14 |
CN-, Cl-, S2O32-, S2-, NH3 |
Drinking water, industrial, food/beverage |
Lead (Pb2+) |
400416 |
Solid state |
Refillable reference, nonreplaceable sensor |
0.02 to 20,700 mg/L |
26 to 29 mV/ decade |
4 to 7 |
Hg2+, Ag+, Cu2+, Fe2+, Cd2+ |
Industrial, environmental |
Nitrate (NO3-) |
400421 |
Polymer/ PVC membrane |
Refillable reference, replaceable sensor modules (1 replacement included) |
0.4 to 62,000 mg/L |
54 to 59 mV/ decade |
2.5 to 11 |
Few, ClO4-, ClO3-, I-, F- |
Environmental, industrial, agriculture, food/ beverage |
Potassium (K+) |
400427 |
Polymer/ PVC membrane |
Refillable reference, replaceable sensor modules (1 replacement included) |
0.04 to 39,000 mg/L |
54 to 59 mV/ decade |
2 to 12 |
Cs+, NH4+, TI+, H+, Ag+, Tris+, Na+, Li+ |
Agriculture, food/beverage, environmental, industrial |
Silver/Sulfide (Ag+/S2-)** |
400433 |
Solid state |
Refillable reference, nonreplaceable sensor |
0.01 to 107,900 mg/L (Silver); 0.003 to 32,100 mg/L (Sulfide) |
54 to 59 mV/decade (Silver); 26 to 29 mV/decade (Sulfide) |
2 to 12 |
Hg2+ |
Industrial, environmental, food/beverage |
Sodium (Na+) |
400439 |
Glass sensor |
Refillable reference, nonreplaceable sensor |
0.1 to 23,000 mg/L |
54 to 59 mV/ decade |
>9 |
H+, K+ |
Industrial, agriculture, power, food/beverage |
• Wide selection of 15 electrodes for 16 different parameters
• 4 sensor technologies - gas sensing, polymer/PVC membrane, solid state, glass sensor
• Combination (i.e. full-cell), refillable electrodes with double-junction reference*
• BNC connection, 3 foot cable
• Made in the U.S.A
• Reference/electrode fill solution, ionic strength adjustor, and small bottle of standard** included with each ISE
• 12 month warranty for solid state, glass sensor, and gas-sensing ISEs; 9 month warranty for ISEs with polymer/PVC membrane
1 unit